美國(guó)CHEMetrics K-0603 水中鋁儀器比色分析包
Aluminum Vacu-vials Kit
Catalog No. K-0603
Range: 0.04-0.25 ppm
Method: Eriochrome Cyanine R (ECR)
Kit comes in a cardboard box and contains everything needed to perform up to 29 tests (except distilled water): thirty ampoules1, Activator Solution, Neutralizer Solution2, 25 mL sample cup, ampoule blank, 1.0 mL syringe, and instructions.
Vacu-vials Kits require the use of the V-2000 Photometer or a spectrophotometer capable of accepting a 13 mm diameter round cell. Instrument sold separately.
1Although the test kit contains 30 ampoules, a fresh reagent ampoule blank must be prepared for each series of tests; therefore, the number of samples that can be tested with each kit will vary from a maximum of 29 to a minimum of 15.
2The Neutralizer Solution is supplied as a dry chemical with NO expiration date. Once reconstituted, it has a shelf-life of 6 weeks that can be extended to 3 months if stored in the refrigerator when not in use.
測(cè)試步驟:

Vacu-vials® for Instrumental Colorimetric Analysis
儀器法測(cè)試包
測(cè)試范圍
|
測(cè)試方法
|
測(cè)試包目錄號(hào)
|
0.04-0.25 ppm
|
鉻花青R
|
K-0603
|
參考:APHA(美國(guó)公共衛(wèi)生協(xié)會(huì))標(biāo)準(zhǔn)方法,第21版,方法3500-AI B(2005);
鉻花青R(ECR)法測(cè)定水中的鋁,Keneth E Shull和Gene R. Guthan, 第1456-1468頁(yè), J.AWWA(美國(guó)自來水協(xié)會(huì)),1967年11月。
鋁在地球表層形成多種礦物質(zhì)。鋁及其合金有很多用途:熱交換器、結(jié)構(gòu)材料、飛行器部件等。鋁(硫酸鋁鉀)用在水處理中作為懸浮顆粒的絮凝劑,但是他的使用可能導(dǎo)致最終飲用水中鋁的濃度升高。飲用水的最大二次污染允許限度為0.05—0.2mg/L。
鋁的Vacu-vials®測(cè)試方法是以鋁和鉻花青R(ECR)之間的反應(yīng)為基礎(chǔ)的,這兩種物質(zhì)在pH值接近6.0時(shí),反應(yīng)形成一種紅色色淀,這種色淀的形成于樣品中的鋁的數(shù)量成比例。測(cè)量結(jié)果用ppm鋁表示。
美國(guó)CHEMetrics中國(guó)代理商:上海益朗儀器有限公司 http://www.shylyq.com